You are here: Metrohm USA >> Support >> FAQs >> Electrodes - Cleaning
Introduction
This protocol covers all combination electrodes and electrode pairs consisting of a glass/pH indicator and a Ag/AgCl reference with 3M KCl salt-bridge isolution.
Instrumentation and reagents
- All pH meters and all titrators that can be operated independently as pH meters may be used
- Standard pH buffered solutions, minimum of two must be used
Procedure
- Microprocessor-based pH meters and titrators
- Calibrate the electrode as per meter/titrator instructions.
- Measure the potentials (mV) of two or more buffer solutions.
- Analog pH meter
- Zero the meter first by shorting the electrode inputs.
- Measure the potentials (mV) of two or more buffer solutions, and compare them with those listed in Table 1.
Observations
|
pH
|
mv (vs. Ag/AgCl - 3M KCl)
(Temperature,°C)
|
| |
20° |
23° |
25° |
| 2.00 |
291.0 |
294.0 |
296.0 |
| 3.00 |
232.8 |
235.2 |
236.8 |
| 4.00 |
174.6 |
176.4 |
177.6 |
| 5.00 |
116.4 |
117.6 |
118.4 |
| 6.00 |
58.2 |
58.8 |
59.2 |
| 7.00 |
0.0 |
0.0 |
0.0 |
| 8.00 |
–58.2 |
–58.8 |
–59.2 |
| 9.00 |
–116.4 |
–117.6 |
–118.4 |
| 10.00 |
–174.6 |
–176.4 |
–177.6 |
| 11.00 |
–232.8 |
–235.2 |
–236.8 |
| 12.00 |
–291.0 |
–294.0 |
–296.0
|
- Electrode calibration : The values of slope, asymmetry voltage, Uas, or asymmetry pH, pHas, reflect the working conditions of a glass/pH electrode. Cleaning becomes necessary whenever one of the conditions below is observed:
- Slope falls outside of 1.025 and 0.975
- Uas is greater than ±15 mV
- pHas falls outside of 6.75 and 7.25
Cleaning and Conditioning
- Reagents
- Alcohol : methyl, ethyl, or isopropyl.
- Ammonium hydroxide, NH4OH, concentrated.
- Mixed acid cleaner : mix concentrated H2SO4 and HNO3 in 2:1 ratio.
- Procedure
- Gently wipe the glass membrane (bulb portion) with alcohol.
- Submerge the lower 1-cm portion in the acid cleaner for a period as listed in Table 2 below.
- Remove electrode and rinse thoroughly with deionized water.
- For combination electrodes only: Soak the lower 1-cm portion in concentrated NH4OH for 15 minutes (longer if the slope falls below 0.950). Remove electrode and rinse thoroughly with deionized water.
- Store all glass/pH electrodes in the electrode filling solution (e.g., 3M, 4M and sat’d. KCl , sat’d KNO3 for at least 10 minutes before use.
Table 2 : Acid cleaning periods
| Electrode condition |
Soaking period |
| For daily cleaning and between titrations in non-aqueous media |
1 – 5 min. |
| After storage for more than 2 weeks. |
10 – 15 min. |
| Uas exceeds ±15 mV and pHas falls outside of 6.70 and 7.20 |
15 min. |
| Reconditioning after prolong use in oily and viscous materials |
0.5 –12 hrs. |
Appendix
The relationship between temperature and theorectical slope, i.e., mV/pH, of glass/pH electrode is described by the equation :
UM = 0.198357 T + 54.20 mV,
where: UM is the theorectical slope, or mv/pH; T is temperature, in °C.
Table 3 : Effect of electrolytes on pH buffer potentials*
| pH |
Potential mV - 3M KCl |
Potential mV - Sat’d KCl |
Potential mV - SCE ** |
| 4.00 |
176.4 |
165.4 |
214.4 |
| 7.00 |
0.0 |
–11.0 |
38.0 |
| 10.00 |
–176.4 |
–187.4 |
–138.4 |
* Measured with a glass combination pH electrode at 25°C.
** SCE = stanard calomel electrode = Hg/Hg2Cl2 - sat’d KCl