Home
Products
Apps
Support
Follow
Company
 
 

Automatic Karl Fischer Water Determination with the 774 Oven Sample Processor

You are here:   Metrohm USA   >>   Support   >>   FAQs   >>   Ovens - Water Determination


 

Based on Metrohm® Application Bulletin 280/1 e

Introduction

In principle, the 774 Oven Sample Processor can be used with all samples that release water when heat is applied. However, the KF oven method is essential whenever direct volumetric or coulometric Karl Fischer titration is impossible because the sample contains interfering components or, owing to its consistency, is difficult to place in the titration vessel.

The combination of the 774 Oven Sample Processor with a Metrohm coulometer is ideal for samples with a low water content. Food, pharmaceutical products, plastics or petrochemical products can be analyzed automatically and extremely accurately. On the other hand, a Metrohm volumetric Karl Fischer titrator is preferred for a sample with a very high water content.

The water is driven out from the heated sample by a stream of dry carrier gas and transferred to the titration vessel, where the water is determined by a KF titration. For temperature-sensitive samples, e.g., food, the water can be released gently at lower temperatures by simultaneous extraction with methanol. In this way, it is possible to prevent any water released by decomposition or any water of crystallization that may split off from being determined and therefore produce incorrect results.

This Application Bulletin uses examples from the food, pharmaceutical, plastics and petrochemical industries to describe automatic Karl Fischer water determination using the 774 Oven Sample Processor and the 756 KF Coulometer. Information applying to the combination of the Oven Sample Processor with a volumetric KF titrator (758 KFD Titrino or 841 Titrando®) is given in brackets.

Instrumentation and accessories

Instrumentation

774 Coulometric System

  • 774 KF Oven Processor (Brinkmann Catalog No. 020816058)
  • 756 KF Coulometer (Brinkmann Catalog No 020815051)
  • Test tool for oven temperature
  • Calibrated reference (Brinkmann Catalog No. 020978708)

774/841 Volumetric System

  • 774 KF Oven Processor (Brinkmann Catalog No. 020816031)
  • 841 KF Titrando (Brinkmann Catalog No. 020840005)
  • Dosino® with 10 ml buret (Brinkmann Catalog No. 020763124)
  • Titrando to PC connecting cable (Brinkmann Catalog No. 020241004)
  • Test tool for oven temperature
  • Calibrated reference (Brinkmann Catalog No. 020978708)

Reagents

  • Hydranal® Coulomat AG Oven, Riedel-de Haën No. 34739 (for coulometric KF titration)
  • Hydranal Methanol Dry, Riedel-de Haën No. 34741 (as additional extraction agent or for volumetric KF titration)
  • Hydranal Composite 5, Riedel-de Haën No. 34805 (for volumetric KF titration)
  • Hydranal Standard sodium tartrate dihydrate, Riedel-de Haën No. 34803
  • Hydranal Water Standard KF Oven (potassium citrate monohydrate), Riedel-de Haën No. 34748
  • Molecular sieve, pore size 0.3 nm, Brinkmann Catalog No. 020689536
  • Carrier gas (air or N2)

Sample preparation

The sample is thoroughly mixed under conditions that are as dry as possible. The sample weight depends on the water content. Sample weights that are too low have a negative effect on the measuring accuracy (weighing error); the maximum sample weight is limited by the reagent capacity (1,000 mg H2O per 100 ml reagent for Hydranal Coulomat AG Oven). The absolute amount of water transferred to the titration vessel for analysis should be in the range 300 µg –1,000 µg. We also recommend that the KF reagent is changed once per week under «normal» operation, more often if necessary.

The appropriate amounts of sample, which should previously have been conditioned for at least 24 h in the surrounding air, are weighed into the sample vessels. The vessels are then sealed with PTFE-coated septa, which have also been conditioned. Gloves should be worn when working to avoid contaminating the samples and prevent contact with skin moisture.

Table 2 clearly displays the relationship between the amount of sample used in the analysis and its water content.

Instrument preparation

On the 774 Oven Sample Processor, the lift positions for the conditioning vessel (rinsing position) and the sample vessels (working position in the oven) as well as the maximum lift path are configured. The previously programmed methods for the Oven Sample Processor and the KF Coulometer (or the KF Titrino) are called up by Tiamo™ titration software or directly on the instruments themselves and, if necessary, modified to suit the samples to be analyzed.

The flow rate of the carrier gas (air or N2) should lie between 40 ml/min–60 ml/min. For solid samples, 40 ml/min is normally used; for liquid samples, 60 ml/min is set. The gas flow must be large enough to transfer the released moisture as quickly as possible to the titration vessel, and small enough to guarantee the complete absorption of the water in the titration solution.

The vessels for conditioning, system preparation, determination of the blank value (see under «Titration sequence») and the sample vessels are placed on the rack of the 774 Oven Sample Processor. For coulometric KF titration, the titration vessel is filled with 150 ml Hydranal Coulomat AG Oven (for volumetric KF titration with approx. 30 ml–40 ml Hydranal Methanol Dry) and then conditioned, i.e., titrated to dryness.

Temperature ramps

For samples whose temperature behavior is unknown, a so-called temperature ramp is run (available temperature range 50°C–250°C). The heating rate should be selected so that it can be guaranteed that the temperature can actually be achieved inside the sample. Figure 1 shows a temperature ramp in which the sample is heated from 50°C to 250°C in 100 min, i.e., the heating rate is 2°C/min.

While a temperature ramp is being recorded, it is possible to record both the amount of water released and the drift as a function of time (see Figs. 5 and 6). By using the heating rate and the elapsed time, it is possible to calculate the associated temperature curve (see Fig. 2). This allows statements about the kinetics of the water release as a function of the temperature to be made. In addition, with thermally unstable samples any decomposition reactions in which water is released can be recognized.
In the temperature or heating up curve of the tartrate sample (see Fig. 2), it can be seen that this releases the sample’s surface water uniformly up to a temperature of approx. 110°C. In parallel the drift sinks from approx. 500 µg/min to 100 µg/min. Between 110°C and 125°C the water of crystallization contained in the sodium tartrate dihydrate is released; this can also be very clearly recognized by the occurrence of a «drift peak». When all the water has been driven off from the sample, the drift sinks to its basic value of approx. 10 µg/min.

The temperature curve is used to determine the optimum oven temperature for driving off the water in the sample. This temperature should be high enough for the water to be extracted completely without any decomposition of the sample occurring; the reaction time should also be kept as short as possible. For sodium tartrate dihydrate, 160°C is recommended as the analysis temperature. The analytical procedure when recording a temperature ramp is the same as for a determination at a given, fixed temperature (see below). Figure 3 shows the temperature curve of a sample of food, which, after releasing all its water, starts to decompose from approx. 180°C.

This temperature curve is used to select an analysis temperature of 120°C for the sample in order to extract all the water it contains. Any transfer of the water of decomposition formed in the titration vessel at higher temperatures (from approx. 180°C) must be avoided at all cost.

Analytical procedure

Each analysis consists of the following steps:

  • Conditioning the titration vessel
  • Extracting the water from the sample
  • Transporting the water to the titration vessel
  • Karl Fischer titration
  • Calculation of the result

 


Fig. 1: Diagram showing a possible temperature ramp.


Fig. 2: Temperature curve of sodium tartrate dihydrate; amount of released water and the associated drift as a function of the temperature.


Fig. 3: Temperature curve of a sample of food that decomposes from 180°C.

Conditioning

Conditioning, i.e., titrating the titration vessel to dryness, is carried out by stirring without increasing the temperature (oven temperature = initial temperature) with the 774 Oven Sample Processor in the so-called conditioning position. This conditioning step must be carried out before every determination. The oven is then heated up to the selected analysis temperature. Only now is the water driven off from the sample by a combination of a dry stream of carrier gas and the increased temperature of the oven and transferred to the KF titration vessel, where the determination by a Karl Fischer titration takes place. This titration may be coulometric (with the 756 KF Coulometer) or—for samples with a high water content (>50%)—volumetric (with the 758 KFD Titrino or 841 Titrando). The latter method should be used if very small sample weights and the associated large measuring inaccuracies are to be avoided.

The double hollow needle of the 774 Oven Sample Processor can be lowered almost to the base of the sample vessel and in this way the carrier gas can be passed through the heated sample. With liquid samples this achieves an additional extraction effect. For the analysis of solid samples, a shortened penetration needle is normally used [headspace technique; this requires the use of special adapters (Brinkmann Catalog No. 020000201)]; this only penetrates the sample vessel a few millimeters deeper than the exhaust needle. In this way turbulent mixing of the sample and contamination of the following samples or the conditioning vessel by substances adhering to the needle tip can be avoided. This also prevents the penetration needle from being bent, as this could happen if it comes into contact with light, voluminous samples, such as granulated plastics.

Water extraction

Carrier gas dried by the molecular sieve is passed through or over the heated sample and drives out the water it contains (see Fig. 4). The temperature of the oven can be varied according to the thermal stability of the sample.

Thermosensitive samples can be gently analyzed at lower temperatures by using Hydranal Methanol Dry as the extraction agent. This is placed in the sample vessel together with the sample. Methanol supports the release of water, particularly during the boiling process in the oven. It is transferred to the titration vessel together with the water but does not interfere with the titration as the KF reagent also contains methanol.

It is generally advisable to set an extraction time that depends on the water content of the sample. This ensures that the released water is completely transferred to the titration vessel. We recommend a time of 3 min to 5 min. With a high water content this time must be extended accordingly or the sample should first be diluted in order to avoid extremely long extraction times (>30 min). If methanol is used as an additional extraction agent, then it is important that a sufficiently long extraction time is selected in order to prevent a premature termination of the determination. In this case we recommend at least 15 min.

Fig. 4: Sample vessel with double hollow needle.

Karl Fischer titration

For a coulometric Karl Fischer titration, a modified KF reagent containing iodide is used. During the determination the iodide is oxidized anodically (generator electrode) to iodine, which then reacts with the water. As this is an absolute method, no titer adjustment is necessary. As in volumetric Karl Fischer titration, the endpoint indication is bi-voltametric using a double Pt-wire electrode (indicator electrode) to which a constant alternating current is applied. This creates a potential difference (voltage) between the two platinum wires. If even very small amounts of free iodine are present, the voltage drops suddenly; this indicates the endpoint of the titration.

In volumetric Karl Fischer titration, the titer of the titrant must first be determined by using a certified water standard. In combination with the 774 Oven Sample Processor, this is done by driving off the water from the «KF Oven Standards» sodium tartrate dihydrate at 160°C or potassium citrate monohydrate at 220°C.

Titration sequence

The water content determined by the gas extraction with subsequent KF titration is made up as follows:

Water contentabsolute = water contentsample + blank value + drift

In conditioning, during which the needle is located in the conditioning vessel, any moisture contained in the system is removed until a constant low drift in the range 10 µg/min–20 µg/min is achieved. The conditioning step must be carried out before each new analysis, and also before system preparation and the determination of the blank value. In combination with Tiamo titration software, the start of the sample series takes place via the <COND> button; this starts conditioning both the Oven Sample Processor and also the KF Coulometer or KF Titrino. If Tiamo software is not used, the analysis is started with the <START> key of the 774 Oven Sample Processor. At this point the KF Coulometer or the KF Titrino must be in the STOP mode. With automatic drift correction, the drift value measured immediately before the determination will be used for determining the water content of the sample (subtracted).

System preparation means that the whole system will be adjusted to the selected conditions. An empty sample vessel will be treated in exactly the same way as the following samples, but the value obtained will not be taken into account. We recommend that this step—which at the same time is used for checking that the analysis system is working properly—is carried out before each new sample series.

Apart from the water in the sample, the sample vessel also contains atmospheric humidity; this makes a blank value determination necessary. If methanol is used as an additional extraction agent then its water content must also be determined, i.e., included in the blank value. A threefold determination of the blank value is normally sufficient. The mean value is stored as a Common Variable and taken into account in the calculation of the water content (subtracted).

The system preparation and the blank value determinations must be carried out under exactly the same conditions as the analysis of the samples. In order to do this, an empty system preparation vessel and three empty (or filled with solvent) blank value vessels are placed on the rack of the 774 Oven Sample Processors and processed before the samples.

Instrument settings

Table 1: Control and titration parameters

  Parameter Setting
Control parameters coulometry: EP at U 50 mV
  Dynamics 70 mV
  Max.rate max. µg/min
  Min.rate 15 µg/min
  Stop crit: rel.drift
  Rel.drift 10 µg/min
  Start drift 10 µg/min
Titration parameters coulometry: I(pol) 10 µA
Preselections coulometry: generator I 400 mA
Control parameters volumetry: EP at U 250 mV
  Dynamics 100 mV
  Max.rate max. ml/min
  Min.volume incr. min. µl
  Stop crit: drift
  Stop drift 20 µl/min
Titration parameters volumetry: Titr.direction
  I(pol) 50 mA
Oven settings: Initial temp. depending on sample, normally 50°C
Gas flow: Flow rate 40 ml/min–60 ml/min
  Gas type air or N2
  Gas flow factor 1.0 (for air or N2)

Troubleshooting

Procedure for poor precision (reproducibility):

  • Optimize the titration and control parameters
  • Check whether the sample vessels are tightly sealed
  • Clean electrodes: Clean indicator electrode mechanically with moist polishing powder (aluminum oxide or toothpaste), rinse and dry; treat with concentrated HNO3 or change the electrode; align Pt-pins parallel; clean generator electrode with concentrated HNO3
  • Drift too high: Check titration cell, septum and/or seals leaks; molecular sieve exhausted; poorly conditioned cell; ensure thorough mixing
  • KF reagent contaminated/used up: Change the solution; use a different batch number if necessary
  • Check electrical contacts
  • Check oven temperature
  • Balance: Too inexact, drafts, temperature influences, temperature equilibrium not reached; sample weight not optimal/too low Possibly carry out a validation of the analysis system

Validation of the analysis system

GLP (Good Laboratory Practice) requires, among other things, that the correctness and precision of analytical instruments is checked at regular intervals by using SOPs (Standard Operating Procedures). We recommend validating the analytical system as a whole by carrying out a series of determinations with certified standards (see Metrohm Application Bulletins No. 255 and 273).

The validation of the 774/756 and 774/758 or 774/841 Systems is carried out by using sodium tartrate dihydrate or potassium citrate monohydrate. These KF standards are supplied with analysis certificates in which their exact water content is given. A tenfold determination is carried out with different sample weights. The relative standard deviation should be less than ≤1.0%.

In addition to the start-up check routines of the instruments, the use of the built-in diagnosis programs and regular service, we also recommend checking the electronic components of the instruments (756 KF Coulometer or 758 KFD Titrino or 841 Titrando) by using the 767 Calibrated Reference.

A measuring insert for the heating block with a built-in temperature sensor is used in combination with a calibrated temperature measuring instrument as an oven temperature test tool for the 774 Oven Sample Processor. The measuring setup is inserted into the oven block in the same way as a sample vessel and fixed in position. When the set temperature has been reached, the inner temperature measured by the test tool is compared with the display of the Oven Sample Processor. If the measured temperature varies greatly from the oven temperature displayed (more than ±4°C), please contact Metrohm Service.

In order to check the gas flow of the Oven Sample Processor, a calibrated flowmeter is included in the gas flow of the system and the flow rate measured by the test tool is compared with the set value. The permissible tolerance of the flowmeter (with microbridge flow sensor) built into the 774 Oven Sample Processor is ±20%. The external flowmeter cannot be purchased from Metrohm AG.

Examples of applications

The following table provides an overview of samples analyzed by the 774 Oven Sample Processor. In each case the Karl Fischer water determination was carried out in combination with the 756 KF Coulometer using N2 as the carrier gas (flow rate 40 ml/min). The water content as well as the sample weights and analysis temperatures are given.

Table 2: Water content of samples from various branches together with their analysis parameters

Sample T (°C) Sample wt. (g) Water content(ppm)
Foodstuffs:      
Lyophilizate 120 0.06 14,000
Aromas1 100 0.03–0.08 25,000–54,000
(Malto-)Dextrin 100 0.03 0.08 47,000–8,8000
Lactose monohydrate 155 0.06 52,000
Skimmed milk powder2 90 0.06 44,000
Whole milk powder2 90 0.06 39,000
Sweet whey powder2 90 0.06 49,000
Glucose monohydrate2 90 0.06 89,000
Maltose monohydrate3 120 0.06 59,000
Roast coffee, ground 145 0.06 48,200
Garlic powder 110 0.2 35,000
Mineral mixtures 110 0.3 60,000–70,000
Plastics:      
Polypropylene 170 3.0 380
Polyethylene4 115 3.0 40
Olefins 180 3.0 100
Polyamide 180 0.3 7,800
Polyoxymethylene (POM) 170 0.3 1,400
Polystyrene5 120 0.05–0.2 200–500
Refinery products:      
Transformer oil 150 3.0 80
Mineral oil 120 1–3 10–100
Insulating oil 140 3 5
Crude oil 140 2 500–1,200
Additive 120 0.01–0.03 44,000
Antimony dialkyldithio-carbamate in crude oil 50–130 3.0 700
Pharmaceutical products:      

Collagens

160

0.07 – 0.4

106,000 – 919,000

Denture cleaner, Effervescent tabletts6 70 0.2 – 1.5 38,000
Drugs 140 0.04 67,000
Lyophilizate 150 0.01 50,000
Vitamin C 80–90 0.1 30,000
Others:      
Sodium tartrate dihydrate 160 0.02–0.08 155,000
Potassium citrate monohydrate 220 0.03 55,500
Methanol 110 0.5 230
Polyammonium compounds 220 0.03–0.3 10,000–850,000
Emulsified fat compound 220 0.03–0.08 850,000
Formamidosulfonic acid 220 0.2–0.3 <10,000
Pigment 100 0.03–0.3 79,000
Polyolether 150 0.6–1.3 1,500
Dibutene 100–140 0.03 250
Inorganic salt 100 0.01–0.03 580
Lithium cobaltite 100 0.45–1.0 64
Building rubble, surface water 50–60 0.3 6,000
Building rubble, bound water 85–140 0.3 10,000
   
1 With the 774 Oven Sample Processor, the dissolution of the sample in KF reagent, which is required in a direct Karl Fischer titration and which frequently cannot be carried out completely, is no longer necessary.
2 With the addition of 4 ml methanol as the extraction agent. Regular checks of the gas flow are necessary as there is a risk that the needle may be blocked.
3 With the addition of 4 ml 1.5-pentandiol, extraction time 600 s.
4 20 ppm water could be determined as the detection limit.
5 With self-expanding plastics, only the surface water can be determined with the 774/756 System. The plastics also contain trapped water. This can only be calculated as the difference between the total water and surface water; the total water is determined by completely dissolving the sample in p-xylene under heating and subsequent volumetric Karl Fischer titration.
6 Above 70°C, the contained carbonate decomposes


Schematic diagram of the 774 Oven Sample Processor and 756 KF Coulometer Analysis System

Method for validating the 774/756 System using sodium tartrate dihydrate (created with 756 KF Coulometer)

  ‘pa      
756 KF Coulometer 5.756.0010
date        1999-12-14                      time         16:27              6
KFC-B                                        774-D j3
parameters      
>control parameters      
  EP at U………………………............................50   mV
  dynamics 70   mV
  max.rate max.   µg/min
  min.rate 15   µg/min
  stop crit: rel.drift    
  rel.drift 5   µg/min
>titration parameters      
  pause 0   s
  extr.time 180   s
  start drift 10   µg/min
  temperature 25.0   °C
  time interval 2   s
  max.titr.time OFF   s
>statistics      
  status: OFF    
>preselections      
  drift corr: auto    
  req.ident: OFF    
  req.smpl size: OFF    
  smpl unit: g    
  limit smpl size: OFF    
  text id1 id1 or C21    
  text id2 id2 or C22    
  text id3 id3 or C23    
  oven: COM1    
  activate pulse: OFF    
============
  ‘fm      
756 KF Coulometer 5.756.0010
date        1999-12-14                      time         16:27              6
KFC-B                                        774-D j3
>calculations      
blank=C39;1;µg    
content=(H2O-C39)*C01/C00/C02;1;ppm    
Water=(H2O-C39)*C03/C00/C04/C05;3;%    
C00= 0.01466    
C01= 1.0    
C02= 1.0    
C03= 100    
C04= 1000    
C05= 1000    
C39= 51.5    
============
  ‘fr      
756 KF Coulometer 5.756.0010
date        1999-12-14                      time         16:26              6
KFC-B                                        774-D j3
smpl size 0.01466   g
smpl heat.time 756   s
sample temp. 160.0   °C
lowest temp. 158.7   °C
highest temp. 160.6   °C
gas flow 38.5   ml/min
drift auto 3.5   µg/min
titr.time 775   s
H2O 2346.9   µg
blank 51.5   µg
content 156575.7   ppm
Water 15.658   %
============

 

The determinations were carried out in 150 ml Hydranal Coulomat AG Oven. The theoretical water content of the standard substance (Apura sodium tartrate dihydrate from Merck®) was 15.65% according to the analysis certificate. Within this validation context, the content was determined as being 15.680 ±0.032% water (srel = 0.21%); this corresponds to a recovery rate of 100.2%.


Fig. 5: Titration curves for the determination of the water content of sodium tartrate dihydrate; released amount of water and drift as a function of time (conditions: T = 160°C, 40 ml/min N2, headspace technique)..

Method for validating the 774/756 System using potassium citrate monohydrate (created with 756 KF Coulometer)

  ‘pa      
756 KF Coulometer 5.756.0010
date        1999-10-28                       time         16:31               11
KFC-B                                        774-D j2
parameters      
>control parameters      
  EP at U………………………............................50   mV
  dynamics 70   mV
  max.rate max.   µg/min
  min.rate 15   µg/min
  stop crit: rel.drift    
  rel.drift 5   µg/min
>titration parameters      
  pause 0   s
  extr.time 180   s
  start drift 10   µg/min
  temperature 25.0   °C
  time interval 2   s
  max.titr.time OFF   s
>statistics      
  status: OFF    
>preselections      
  drift corr: auto    
  req.ident: OFF    
  req.smpl size: OFF    
  smpl unit: g    
  limit smpl size: OFF    
  text id1 id1 or C21    
  text id2 id2 or C22    
  text id3 id3 or C23    
  oven: COM1    
  activate pulse: OFF    
============
  ‘fm      
756 KF Coulometer 5.756.0010
date        1999-10-28                       time         16:31               11
KFC-B                                        774-D j2
>calculations      
blank=C39;1;µg    
content=(H2O-C39)*C01/C00/C02;1;ppm    
Water=(H2O-C39)*C03/C00/C04/C05;3;%    
C00= 0.02588    
C01= 1.0    
C02= 1.0    
C03= 100    
C04= 1000    
C05= 1000    
C39= 74.4    
============
  ‘fr      
756 KF Coulometer 5.756.0010
date        1999-10-28                       time         16:30              11
KFC-B                                        774-D j2
smpl size 0.02588   g
smpl heat.time 592   s
sample temp. 220.0   °C
lowest temp. 218.1   °C
highest temp. 220.3   °C
gas flow 39.2   ml/min
drift auto 4.6   µg/min
titr.time 613   s
H2O 1520.9   µg
blank 74.4   µg
content 55892.6   ppm
Water 5.589   %
============

 

The determinations were carried out in 100 ml Hydranal Coulomat AG Oven plus 50 ml methanol. The theoretical water content of the standard substance (Hydranal Water Standard KF Oven from Riedel-de Haën) was 5.59% according to the analysis certificate. Within this validation context, the content was determined as being 5.589 ± 0.049% water (srel = 0.88%); this corresponds to a recovery rate of 100.0%.


Fig. 6: Titration curves for the determination of the water content of potassium citrate monohydrate; released amount of water and drift as a function of time (conditions: T = 220°C, 40 ml/min N2, headspace technique).