You are here: Metrohm USA >> Support >> FAQs >> Titration - Stopcock
- Close Tiamo software then re-start. All dosing units are initialized and sometimes that is enough to clear the message.
- If message persists, remove Dosino® from the buret then unscrew the buret from the reagent bottle. Follow the instructions below to dismantle the unit.
- If the Dosino drive shaft is extended into the buret, unscrew the buret and Dosino together from the reagent bottle. Pull the buret cylinder with the black valve disk down and away from the drive. Using the manual control, activate “fill” to pull the drive shaft up. Follow the directions below to dismantle and clean the unit.
Cleaning
- Liquid detergent, e.g., ethanol
- Diluted soap water
- Rinse with distilled water
- Dry with fiber-free cloth
- Ultrasonic bath
- NO abrasives!
| |
|
No damage to the surface |
 |
 |
 |
 |
 |
 |
| • centering tube |
• housing |
• cylinder |
• piston peg |
• black valve disk |
• white distribution disk |
Follow the steps below to re-assemble the unit.
Assembly of glass cylinder
- Place glass cylinder on a plane surface.
- Position first ridge of the piston peg into the glass cylinder.
- Carefully place the centering tube over the piston peg. Match the width of the tongues at the bottom.
- Place the housing over the centering tube.
- ONLY now apply pressure on the top to merge all parts together.